Application of 25MHz B-Scan Ultrasonography to Determine the Integrity of the Posterior Capsule in Posterior Polar Cataract

8Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Purpose. To report the application of 25MHz B-scan ultrasonography (MHzB) to determine the integrity of the posterior capsule (PC) in posterior polar cataract (PPC). Methods. Patients with whom PPC was clinically diagnosed using slit lamp microscopy who underwent 25 MHzB before phacoemulsification were retrospectively reviewed. The status of the PC was determined by 25 MHzB before phacoemulsification and confirmed during cataract surgery. Results. In total, 21 eyes in 14 clinically diagnosed PPC patients were enrolled in this study. Out of 25MHzB images, 19 PCs were found to be intact, while 2 showed dehiscence before cataract surgery. During phacoemulsification, 17 PCs were observed to be intact, while 4 PCs showed posterior capsule rupture (PCR). These 4 PCR cases included the above 2 eyes, in which preexisting dehiscence was detected by 25 MHzB. The other 2 PCR cases showed high reflectivity between high echoes in posterior opacities and the PC, indicating synechia between the PPC and PC. Conclusion. This is the first report to show that 25MHzB can be used to clearly visualize the status of the PC in PPC. These results, in turn, could be used to select the appropriate treatment and to thereby avoid further complications during PPC surgery.

Cite

CITATION STYLE

APA

Guo, Y., Lu, C., Wu, B., Gao, J., Li, J., Yuan, X., & Tang, X. (2018). Application of 25MHz B-Scan Ultrasonography to Determine the Integrity of the Posterior Capsule in Posterior Polar Cataract. Journal of Ophthalmology, 2018. https://doi.org/10.1155/2018/9635289

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free