The measurement of hotspot electron temperature is a paramount technique of implosion physics research in inertial confinement fusion. This study proposes a novel quasi-coaxis dual-energy flat spectral response high-resolution X-ray imaging instrument comprising a dual-channel total-reflection Kirkpatrick-Baez microscope and two flat non-periodic multilayer mirrors, which can image at 6.4 ± 0.5 and 9.67 ± 0.5 keV simultaneously. Various theoretical simulations were performed to verify the performance and feasibility of the imaging instrument, which was assembled and characterized in a laboratory. Experimental results show that the imaging instrument could achieve a high spatial resolution of 5 µm in a ± 150 µm field of view (FOV), the root mean square(RMS) deviation values of the measured reflection efficiency are 1.71% and 1.82% for the 6.4 keV and 9.67 keV imaging channels, respectively, in the ± 150 µm FOV.
CITATION STYLE
Li, W., Mu, B., Ren, K., Xu, J., Chen, L., Li, M., … Wang, F. (2022). Development of a quasi-coaxis dual-energy flat spectral response X-ray imaging instrument for measuring hotspot electron temperature. Optics Express, 30(6), 8777. https://doi.org/10.1364/oe.454237