Abstract
A sensitive and selective procedure is presented for the voltammetric determination of cobalt. The procedure involves an adsorptive accumulation of cobalt Pyrogallol Red (PGR) complex on a stationary mercury drop electrode, followed by cathodic stripping voltammetry measurement of the reduction current of the adsorbed complex at -1.17 V (vs. Ag/AgCl). The optimum conditions for the determination of cobalt include pH 11.0, 35 μM Pyrogallol Red, an accumulation potential of -0.9 V (vs. Ag/AgCl) and scan rate 80 mV s -1. The peak current is proportional to the concentration of cobalt over the concentration range 5.0 to 280 ng mL-1 with a detection limit of 1 ng mL-1 and an accumulation time of 140 s. The method was applied to the determination of cobalt in analytical grade NaCl and water samples. Copyright (C)2013 SCS.
Author supplied keywords
Cite
CITATION STYLE
Hasanpour, F., Teimori, H., Fouladgar, M., & Taei, M. (2013). Trace and selective determination of cobalt(II) in water and salt samples using cathodic adsorptive stripping voltammetry in the presence of Pyrogallol Red. Journal of the Serbian Chemical Society, 78(5), 717–724. https://doi.org/10.2298/JSC120305070H
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.