Abstract
Introduction: Lipohypertrophy (LH) is caused by repetitively injecting insulin into the same location. This can lead to unpredictable insulin absorption and increased glucose variability (GV). A new medical device, ROTO Track, automatically guides the user to rotate abdominal insulin injections to avoid LH lesions. This study aimed to test whether the medical device could reduce the number of insulin injections in the same subcutaneous area as compared with non-aided standard insulin injection techniques. Methods: In this proof-of-concept cross-over study, baseline data about injection site in the abdominal region were collected blinded for 1 week with a nonguiding version of the device and compared to 1 and 12 weeks of device guidance in 35 people with type 1 diabetes. The device registered time and location of abdominal injections. The primary endpoint was a “rotation score.” Secondary endpoints included number and size of LH, GV, and hemoglobin A1c. Results: The rotation score improved significantly from a baseline mean of 40.2% to 49.9% after 1 week (confidence interval: 2.2-17.2%, P =.012) and improved further after 12 weeks to 52.2% (P
Author supplied keywords
Cite
CITATION STYLE
Klarskov, C. K., Hamid, Y. H., Tjalk-Bøggild, R., Tarnow, L., & Kristensen, P. L. (2020). A New Medical Device for Improved Rotation of Insulin Injections in Type 1 Diabetes Mellitus: A Proof-of-Concept Study. Journal of Diabetes Science and Technology, 15(5), 1111–1120. https://doi.org/10.1177/1932296820950688
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.