Abstract
PURPOSE: To assess the performance of a new urinary intermittent catheter (IC) prototype designed with a micro-hole drainage zone compared to a conventional eyelet catheter (CEC) in terms of flow-stop, bladder emptying, and hematuria. DESIGN: Randomized controlled crossover studies. SUBJECT AND SETTING: The sample comprised 15 male healthy volunteers (HV) and 15 IC users, along with 15 female HV and 15 IC users. The age range was lower for HV participants than for IC users (range: 20-57 years for HV vs 21-82 years for IC users). The study setting was the Department of Urology, located in Rigshospitalet, Copenhagen. METHODS: Number of flow-stop incidents, residual urine volume at first flow-stop (RV1), and dipstick hematuria were measured during and after catheterization by a health care professional (HV) and by self-catheterisation (IC-users). Results from the 3 studies were combined for HV and IC users on RV1 and number of flow-stop incidents but separated on sex. For incidents of hematuria, an effect of underlying condition was assumed, and a combined analysis on sex was performed, separating HV and IC users. RESULTS: When compared to the micro-hole drainage zone design, catheterizations with CEC resulted in a significantly higher mean RV1 (mean difference: 49 mL in males and 32 mL in females, both P
Author supplied keywords
Cite
CITATION STYLE
Landauro, M. H., Tentor, F., Pedersen, T., Jacobsen, L., & Bagi, P. (2023). Improved Performance With the Micro-Hole Zone Intermittent Catheter: A Combined Analysis of 3 Randomized Controlled Studies Comparing the New Catheter Technology With a Conventional Eyelet Catheter. Journal of Wound, Ostomy and Continence Nursing, 50(6), 504–511. https://doi.org/10.1097/WON.0000000000001029
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.