Abstract
Total hip arthroplasty (THA) has in recent years trended toward a younger, more physically demanding patient population. Mid- to long-term studies of all ages of THA patients using highly cross-linked polyethylene (HXLPE) have been favorable, but concerns about its long-term failure and wear-related complications remain for young THA patients. In this narrative review, a search of the PubMed/MEDLINE and Cochrane databases was performed, and we identified six studies with a minimum 15-year follow-up of HXLPE with various femoral head materials. Wear-related revisions were exceedingly low for patients under the age of 55, with variable reports of non-clinically significant osteolysis. Higher activity levels, a larger femoral head size, and femoral head material were not associated with greater long-term wear rates. Young THA with metal or ceramic on HXLPE is exceedingly durable with favorable outcomes at follow-ups of over 15 years.
Author supplied keywords
Cite
CITATION STYLE
Deans, C. F., Buckner, B. C., & Garvin, K. L. (2023, October 1). Wear, Osteolysis, and Aseptic Loosening Following Total Hip Arthroplasty in Young Patients with Highly Cross-Linked Polyethylene: A Review of Studies with a Follow-Up of over 15 Years. Journal of Clinical Medicine. Multidisciplinary Digital Publishing Institute (MDPI). https://doi.org/10.3390/jcm12206615