Abstract
The noninvasive measurement of infant lung function during unsedated sleep in infants has been a long-standing objective in paediatric respiratory medicine. This note reports on the design and performance of a head-out volume-displacement plethysmograph (VDP) that overcomes some of the limitations of traditional lung function apparatus. The VDP comprises a rigid acrylic box with an integral water-sealed spirometer and a novel neck seal. The bilayer neck seal is of variable compliance and is comfortable and simple to use. The spirometer permits volume resolution of 1.5 mL and a dynamic range in excess of 100 mL. The frequency response extends from 0-7 Hz. Spirometer inertance was measured as 0.0015 kPa·L-1·s-2, resistance 0.021 kPa·L-1·s-1 and box capacitance 0.18 L·kPa-1. Tidal volume, respiratory rate and changes in functional residual capacity can be recorded during unsedated rapid eye movement and nonrapid eye movement whilst monitoring with conventional polysomnographic methods. The head-out configuration allows additional instrumentation to be implemented with ease, avoids facial stimulation and allows unimpeded access to the upper airway. A polysomnograph illustrating the limitations of respiratory inductance plethysmography signals and typical changes in functional residual capacity are shown.
Author supplied keywords
Cite
CITATION STYLE
Wilson, S. J., O’Brien, C., Harris, M. A., & Masters, I. B. (1998). Measuring tidal volume and functional residual capacity change in sleeping infants using a volume displacement plethysmograph. European Respiratory Journal, 12(5), 1186–1190. https://doi.org/10.1183/09031936.98.12051186