Combined fluorescence lifetime and surface topographical imaging of biological tissue

  • Hopkinson C
  • Matheson A
  • Finlayson N
  • et al.
1Citations
Citations of this article
9Readers
Mendeley users who have this article in their library.
Get full text

Abstract

In this work a combined fluorescence lifetime and surface topographical imaging system is demonstrated. Based around a 126 × 192 time resolved single photon avalanche diode (SPAD) array operating in time correlated single-photon counting (TCSPC) mode, both the fluorescence lifetime and time of flight (ToF) can be calculated on a pixel by pixel basis. Initial tests on fluorescent samples show it is able to provide 4 mm resolution in distance and 0.4 ns resolution in lifetime. This combined modality has potential biomedical applications such as surgical guidance, endoscopy, and diagnostic imaging. The system is demonstrated on both ovine and human pulmonary tissue samples, where it offers excellent fluorescence lifetime contrast whilst also giving a measure of the distance to the sample surface.

Cite

CITATION STYLE

APA

Hopkinson, C., Matheson, A. B., Finlayson, N., Tanner, M. G., Akram, A. R., & Henderson, R. K. (2024). Combined fluorescence lifetime and surface topographical imaging of biological tissue. Biomedical Optics Express, 15(1), 212. https://doi.org/10.1364/boe.504309