Evaluating the analytical performance of five new coagulation assays for the measurement of prothrombin time and activated thromboplastin time

18Citations
Citations of this article
28Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Introduction: New methods for coagulation tests require careful assessment before routine use. We evaluated the analytical performance of five new coagulation assays for measuring prothrombin time (PT) and activated partial thromboplastin time (aPTT). Methods: PT Rec, PT Owren, aPTT, aPTT Lupus and aPTT Screen assays (Roche Diagnostics) were evaluated on cobas t 711 and cobas t 511 analysers (Roche Diagnostics) at four European centres. Analytical performance and method comparisons with relevant commercially available assays were performed to Clinical Laboratory Standards Institute guidelines using residual anonymized samples. Lot-to-lot comparison and equivalency of the cobas t analysers were also assessed; reference ranges were determined using samples from apparently healthy volunteers. Results: Overall, coefficients of variation were ≤1.3% for within-run precision and ≤6.3% for total reproducibility across all sites. Deming regression analyses showed good agreement between each assay (cobas t 711) and respective comparator method (Pearson's r: 0.964-0.999, n > 120 samples/assay/site). Passing–Bablok regression analyses demonstrated equivalence of the two cobas t platforms for use with each assay (Pearson's r ≥ 0.995). Lot-to-lot consistency was high for all assays and comparisons (Pearson's r ≥ 0.998). Reference ranges (2.5th-97.5th percentiles; n = 200 samples/assay) in seconds were 8.4-10.6 (PT Rec), 18.2-27.2 (PT Owren), 23.6-30.6 (aPTT), 24.1-31.7 (aPTT Lupus) and 23.9-33.2 (aPTT Screen). Conclusion: Based on the excellent analytical performance and good agreement with relevant comparator methods, the five coagulation assays on the novel cobas t 711 and cobas t 511 analysers are suitable for routine use in core laboratories.

Cite

CITATION STYLE

APA

Kitchen, S., Geisen, U., Kappelmayer, J., Quehenberger, P., Drieß, J., Lowe, A., … Rozsnyai, G. (2018). Evaluating the analytical performance of five new coagulation assays for the measurement of prothrombin time and activated thromboplastin time. International Journal of Laboratory Hematology, 40(6), 645–654. https://doi.org/10.1111/ijlh.12897

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free