Purpose: The aim of this study was to compare the durability and retention of 4 types of attachments placed over computer aided design/computer aided manufacturing (CAD/CAM) titanium bars when subjected to different pH conditions. Materials and methods: Four commercially available attachments were investigated: Hader Yellow, Hader Red, Ackerman Gold and Ackerman Stainless Steel. These attachments and Ackerman CAD/CAM titanium bars were placed in 2 vessels containing different artificial saliva solutions (pH 7/ pH 4) at 37°C for one month to simulate corrosion conditions, and they were then subjected to mechanical testing (5400 cycles of insertion and removal). Results: The results revealed that there were significant differences in the average values of insertion/removal force due to the pH (F (1, 24)=9.207, P < .05) and the type of attachment (F (3, 24)=11.742, P < .05). Conclusion. More acidic pH values were found to have a negative influence on the retention capacity of the attachments.

Cite

CITATION STYLE

APA

Silva, A. S., Aroso, C., Ustrell, R., Braga, A. C., Mendes, J. M., & Escuin, T. (2015). The influence of saliva pH value on the retention and durability of bar-clip attachments. Journal of Advanced Prosthodontics, 7(1), 32–38. https://doi.org/10.4047/jap.2015.7.1.32