Abstract
Augmentation index (AIx) and subendocardial viability ratio (SEVR) are widely accepted indices of wave reflection and myocardial oxygen demand relative to supply. This study aimed to validate a new tonometric device (IIM-2010A) for obtaining AIx and SEVR from radial artery. A total of 68 outpatients (32 men and 36 women) aged 20 to 76 years (44.7±16.6 years) recruited from a health screening center participated in the study. AIx was obtained from radial pressure using the HEM-9000AI and IIM-2010A devices, while SEVR was measured from carotid pressure with the tonometric method and from radial pressure by the IIM-2010A device. In a subgroup of 24 patients, the measurements of AIx and SEVR were repeated after an interval of 10 minutes. The correlation of radial AIx between the IIM-2010A and HEM-9000AI devices was highly significant (r=0.956, P
Cite
CITATION STYLE
Wang, J. Z., Zhang, Y. L., Hu, F. S., He, Z. J., Yang, X. J., Ma, Z. C., & Sun, Y. N. (2014). A new tonometric device for radial augmentation index and subendocardial viability ratio: Potential use in health screening. Journal of Clinical Hypertension, 16(10), 707–712. https://doi.org/10.1111/jch.12396