Performance of the Packard Tri-Carb® 2770TR/SL liquid scintillation analyzer for 14C dating

2Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

Abstract

We present results that demonstrate the potential of the Packard Tri-Carb® Model 2770TR/SL for radiocarbon dating. For 2 g of sample benzene, a stable background count rate of 0.307 cpm and a stable counting efficiency of 64.78% were determined using a 13-75 keV counting window. Changes to the mathematical routines for t-SIE (quench indicating parameter) calculation and a reduction in the activity of the external standard have enabled stability of the t-SIE to be achieved, and combined with the use of a suitable balance point counting window; all of these factors give the stability of performance required for 14C dating. Calculations based on the above parameters indicate that the limit of detection for 2 g samples, counted for 5000 min, is >48,900 yr BP. The great advantage of this system is that these data were acquired using inexpensive standard 7-mL low potassium borosilicate glass vials. Vial holders manufactured from BGO reduced the background to 0.15 cpm with a minimum effect on efficiency (64.46% for 13-75 keV). A similar calculation of the limit of detection gave >51,700 BP. The use of the BGO vial holders in other instruments employing time-resolved liquid scintillation counting (TR-LSC) (Models 2250CA and 2260XL) also brought about significant improvements in detection limits.

Cite

CITATION STYLE

APA

Passo, C. J., Anderson, R., Roberts, D., & Cook, G. T. (1998). Performance of the Packard Tri-Carb® 2770TR/SL liquid scintillation analyzer for 14C dating. Radiocarbon, 40(1), 193–200. https://doi.org/10.1017/s003382220001804x

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free