2D and 3D dose verification at The Netherlands Cancer Institute-Antoni van Leeuwenhoek Hospital using EPIDs

8Citations
Citations of this article
30Readers
Mendeley users who have this article in their library.

Abstract

A review is given of the clinical use of EPID dosimetry in the Department of Radiation Oncology of The Netherlands Cancer Institute-Antoni van Leeuwenhoek Hospital. All curative plans (almost all IMRT or VMAT) are verified with EPID dosimetry, mostly in vivo. The 2D approach for IMRT verification and the 3D method for VMAT verification are elucidated and their clinical implementation described. It has been shown that EPID dosimetry plays an important role in the total chain of verification procedures that are implemented in our department. It provides a safety net for advanced treatments such as IMRT and VMAT, as well as a full account of the dose delivered. © 2010 IOP Publishing Ltd.

Cite

CITATION STYLE

APA

Mijnheer, B., Mans, A., Olaciregui-Ruiz, I., Sonke, J. J., Tielenburg, R., Van Herk, M., … Stroom, J. (2010). 2D and 3D dose verification at The Netherlands Cancer Institute-Antoni van Leeuwenhoek Hospital using EPIDs. In Journal of Physics: Conference Series (Vol. 250, pp. 97–102). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/250/1/012020

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free