Abstract
This study describes the development of an immunosensory label-free quantification methodology based on surface plasmon resonance (SPR) and its applicability in measuring/ evaluating therapeutic drug monitoring (TDM) of anti-TNF-α monoclonal antibody (adalimumab) in rheumatoid arthritis (RA) patients. The experimental parameters evaluated in this study were immobilising ligands by pre-concentration assays, sensor surface regeneration, ascertaining the method’s sensitivity and correlating the results from quantifying plasma samples by ELISA immunoassay. The results showed that TNF-α quantification values (in RU) were significantly different when comparing patients (~50-250 RU) to controls (~10-20 RU). Likewise, there was 0.97 correlation for patients and 0.91 for healthy volunteers using SPR and ELISA comparison methodologies. SPR immunosensory detection provided a precise, sensitive strategy, along with real-time determination, for quantifying adalimumab, having great potential for clinical routine regarding TDM.
Author supplied keywords
Cite
CITATION STYLE
Bustos, R. H., Zapata, C., Esteban, E., García, J. C., Jáuregui, E., & Jaimes, D. (2018). Label-free quantification of anti-TNF-α in patients treated with adalimumab using an optical biosensor. Sensors (Switzerland), 18(3). https://doi.org/10.3390/s18030691