Comparison of a new non-contact tonometer with Goldmann applanation

60Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A comparison of a new non-contact tonometer (Keeler PulsairTM: Keeler UK Ltd) with the Goldmann applanation tonometer is reported. Measurements of intraocular pressure were obtained from 182 eyes of 94 patients. At low pressures (<10 mmHg) the candidate tonometer tended to overestimate pressures obtained with the Goldmann tonometer whilst at high pressures (>19 mmHg) those obtained by Goldmann applanation were underestimated. Between 10 and 19 mmHg there was no significant difference between readings obtained with either method. Up to 71% of averaged Pulsair measurements fell within ± 3 mmHg of those obtained with the Goldmann tonometer increasing to 78% if pressures ≥30 mmHg obtained with the Pulsair tonometer were excluded. Adopting a screening criterion of ≥21 mmHg (Goldmann) resulted in a sensitivity of 85% and a specificity of 95%. Some evidence that serial Pulsair readings are influenced by the ocular pulse is presented. It is concluded that the Pulsair tonometer can provide clinically useful measurements of intraocular pressure. © 1989, College of Ophthalmologists. All right reserved.

Cite

CITATION STYLE

APA

Moseley, M. J., Evans, N. M., Fielder, A. R., & Hitchings, R. A. (1989). Comparison of a new non-contact tonometer with Goldmann applanation. Eye (Basingstoke), 3(3), 332–337. https://doi.org/10.1038/eye.1989.48

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free