Finite analysis of stability between modified articular fusion technique, posterior lumbar interbody fusion and posteriorlateral lumbar fusion

9Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: It is not clear whether modified facet fusion (MFF) is biomechanically different from traditional fusion techniques such as posterior lateral lumbar fusion (PLF) and posterior lumbar interbody fusion (PLIF). Methods: In this study, a healthy adult Chinese male volunteer was selected to perform 3D reconstruction of CT image data and simulate the successful fusion of L4–5 MFF, PLF and PLIF, respectively. The motion range of L4–5 segments of the model was simulated under 6 working conditions, including forward flexion, extension, lateral flexion and rotation under normal physiological conditions, and the stability of the three fusion procedures in the pathological segments of the lumbar spine was compared. Results: There was no difference in range of motion between MFF model and PLF or PLIF model (P < 0.05). Also, the stiffness of the PLFand the MFF model were comparable (P > 0.05), but were smaller than the PLIF model (P < 0.05). Conclusions: MFF provides reliable stability at the lumbar fixation fusion level and does not differ significantly from PLF and PLIF in terms of range of motion.

Cite

CITATION STYLE

APA

Han, X., Chen, X., Li, K., Li, Z., & Li, S. (2021). Finite analysis of stability between modified articular fusion technique, posterior lumbar interbody fusion and posteriorlateral lumbar fusion. BMC Musculoskeletal Disorders, 22(1). https://doi.org/10.1186/s12891-021-04899-x

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free