Efficient detection of bilirubin in human serum through a displacement approach

6Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

Abstract

The present study provides, for the first time, a method for the fluorescence “Turn ON” quantification of bilirubin (BR) in human serum through a displacement approach. The fluorescent probe THYQ exhibited a 55-fold increase in fluorescence intensity at 555 nm with HSA. The green fluorescent THYQ-HSA complex was highly selective towards bilirubin (BR) (subdomain IB) and exhibited >90% fluorescence quenching (log β = 4.83 ± 0.12) through the displacement of THYQ, whereas the subdomain IIA and subdomain IIIA selective drugs warfarin and ibuprofen did not displace THYQ. The non-fluorescent BR-HSA complex displayed a linear increase in fluorescence with THYQ (log β = 4.81 ± 0.13) to form THYQ-HSA complex. The slope of titration of BR-HSA with THYQ was linearly dependent on [bilirubin] in HSA, thus enabling the fluorescence-based detection of bilirubin below normal (0.7 mg dL−1, 12 μM) to hyperbilirubinemia conditions (12.6 mg dL−1, 216 μM), with the results in good agreement with the clinical findings. The LOD for the detection of BR was 0.004 mg dL−1 (68 nM).

Cite

CITATION STYLE

APA

Singla, N., Ahmad, M., Mahajan, V., Singh, P., & Kumar, S. (2023). Efficient detection of bilirubin in human serum through a displacement approach. Sensors and Diagnostics, 2(6), 1574–1584. https://doi.org/10.1039/d3sd00157a