Beyond Poiseuille: Preservation Fluid Flow in an Experimental Model

  • Singh S
  • Randle L
  • Callaghan P
  • et al.
N/ACitations
Citations of this article
28Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Poiseuille’s equation describes the relationship between fluid viscosity, pressure, tubing diameter, and flow, yet it is not known if cold organ perfusion systems follow this equation. We investigated these relationships in an ex vivo model and aimed to offer some rationale for equipment selection. Increasing the cannula size from 14 to 20 Fr increased flow rate by a mean (SD) of 13 (12)%. Marshall’s hyperosmolar citrate was three times less viscous than UW solution, but flows were only 45% faster. Doubling the bag pressure led to a mean (SD) flow rate increase of only 19 (13)%, not twice the rate. When external pressure devices were used, 100 mmHg of continuous pressure increased flow by a mean (SD) of 43 (17)% when compared to the same pressure applied initially only. Poiseuille’s equation was not followed; this is most likely due to “slipping” of preservation fluid within the plastic tubing. Cannula size made little difference over the ranges examined; flows are primarily determined by bag pressure and fluid viscosity. External infusor devices require continuous pressurisation to deliver high flow. Future studies examining the impact of perfusion variables on graft outcomes should include detailed equipment descriptions.

Cite

CITATION STYLE

APA

Singh, S., Randle, L. V., Callaghan, P. T., Watson, C. J. E., & Callaghan, C. J. (2013). Beyond Poiseuille: Preservation Fluid Flow in an Experimental Model. Journal of Transplantation, 2013, 1–6. https://doi.org/10.1155/2013/605326

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free