Abstract

To assess how well results by different methods for urinary oxalate determinations agree with each other in a clinical setting, we compared six different assays: Hodgkinson and Williams, enzymatic, modified Hodgkinson and Williams, gas chromatography, ion chromatography, and 'high-pressure' liquid chromatography. For the entire group of samples, the mean value by each method agreed relatively closely, although the enzymatic procedure produced a somewhat higher value. All six methods had large coefficients of variation within (8-58%) and between (15-88%) assays. In addition, analytical recovery by most assays was more than 100% of the added oxalate. Analytical recovery of 10 μg of oxalate added per milliliter of urine specimen ranged from 86 to 237%; for 20 μg/mL it was 83 to 320%. Thus for the six methods evaluated, no single method appeared to be superior to the others.

Cite

CITATION STYLE

APA

Zerwekh, J. E., Drake, E., Gregory, J., Griffith, D., Hofmann, A. F., Menon, M., & Pak, C. Y. (1983). Assay of urinary oxalate: Six methodologies compared. Clinical Chemistry, 29(11), 1977–1980. https://doi.org/10.1093/clinchem/29.11.1977