Reproducibility of intra-operative measurement of the mechanical axes of the lower limb during total knee replacement with a non-image-based navigation system

  • Jenny J
  • Boeri C
  • Picard F
  • et al.
N/ACitations
Citations of this article
17Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

OBJECTIVE: The restoration of a normal mechanical axis of the lower limb following total knee prosthesis (TKP) depends on the accuracy of the intra-operative measurement of the femoro-tibial angle. We have studied the reproducibility of intra-operative measurement of the coronal mechanical femoro-tibial axis with the OrthoPilot (Aesculap, Tuttlingen, Germany) non-image-based navigation system. MATERIAL AND METHODS: A consecutive series of 20 TKP (Aesculap SEARCH Evolution prosthesis) implanted by the same surgical team of two senior orthopedic surgeons was analyzed. They used a non-image-based navigation system that allows the mechanical axes of the femur and tibia to be defined with a kinematic analysis. The operating surgeon and assistant surgeon performed the kinematic analysis twice and once, respectively, and measured coronal mechanical femoro-tibial angles in maximal extension and at 90 degrees flexion without varus or valgus stress. RESULTS: The mean intra-observer variation in the measurement of the coronal mechanical femoro-tibial angle in maximal extension was 0.1 degrees (SD = 0.7 degrees). The mean intra-observer variation in the measurement of the coronal mechanical femoro-tibial angle at 90 degrees of knee flexion was 0.2 degrees (SD = 0.6 degrees). The mean inter-observer variation in the measurement of the coronal mechanical femoro-tibial angle in maximal extension was 0.1 degrees (SD = 0.7 degrees). The mean inter-observer variation in the measurement of the coronal mechanical femoro-tibial angle in maximal extension was 0.0 degrees (SD = 0.6 degrees). There were no significant differences and a high correlation between all paired intra- and inter-observer measurements. CONCLUSION: This system allows high reproducibility of the intra-operative measurement of the mechanical axes of the lower limb by a non-image-based kinematic registration of the hip, knee and ankle centers.

Cite

CITATION STYLE

APA

Jenny, J., Boeri, C., Picard, F., & Leitner, F. (2004). Reproducibility of intra-operative measurement of the mechanical axes of the lower limb during total knee replacement with a non-image-based navigation system. Computer Aided Surgery, 9(4), 161–165. https://doi.org/10.3109/10929080500095517

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free