Abstract
Purpose: To investigate the robustness of constrained and simultaneous intravoxel incoherent motion (IVIM) fitting methods and the estimated IVIM parameters (D, D* and f) for applications in brain and low-perfused tissues. Materials and Methods: Model data simulations relevant to brain and low-perfused tumor tissues were computed to assess the accuracy, relative bias, and reproducibility (CV%) of the fitting methods in estimating the IVIM parameters. The simulations were performed at a series of signal-to-noise ratio (SNR) levels to assess the influence of noise on the fitting. Results: The estimated IVIM parameters from model simulations were found significantly different (P < 0.05) using simultaneous and constrained fitting methods at low SNR. Higher accuracy and reproducibility were achieved with the constrained fitting method. Using this method, the mean error (%) for the estimated IVIM parameters at a clinically relevant SNR = 40 were D 0.35, D* 41.0 and f 4.55 for the tumor model and D 1.87, D* 2.48, and f 7.49 for the gray matter model. The most robust parameters were the IVIM-D and IVIM-f. The IVIM-D* was increasingly overestimated at low perfusion. Conclusion: A constrained IVIM fitting method provides more accurate and reproducible IVIM parameters in low-perfused tissue compared with simultaneous fitting. Level of Evidence: 3. J. MAGN. RESON. IMAGING 2017;45:1325–1334.
Author supplied keywords
Cite
CITATION STYLE
Meeus, E. M., Novak, J., Withey, S. B., Zarinabad, N., Dehghani, H., & Peet, A. C. (2017). Evaluation of intravoxel incoherent motion fitting methods in low-perfused tissue. Journal of Magnetic Resonance Imaging, 45(5), 1325–1334. https://doi.org/10.1002/jmri.25411
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.