Detection of a chemical warfare agent simulant in various aerosol matrixes by ion mobility time-of-flight mass spectrometry

216Citations
Citations of this article
97Readers
Mendeley users who have this article in their library.
Get full text

Abstract

For the first time, a traditional radioactive nickel (63Ni) β emission ionization source for ion mobility spectrometry was employed with an atmospheric pressure ion mobility orthogonal reflector time-of-flight mass spectrometer (IM-(tof)MS) to detect a chemical warfare agent (CWA) simulant from aerosol samples. Aerosol-phase sampling employed a quartz cyclonic chamber for sample introduction. The simulant reference material, which closely mimicked the characteristic chemical structure of CWAs as defined and described by Schedule 1, 2, or 3 of the Chemical Warfare Convention treaty verification, was used in this study. An overall elevation in arbitrary signal intensity of ∼1.0 orders of magnitude was obtained by the progressive increase of the thermal AP-IMS temperature from 75 to 275 ∼C. A mixture of one G-type nerve simulant (dimethyl methylphosphonate (DMMP)) in four (water, kerosene, gasoline, diesel) matrixes was found in each case (AP-IMS temperature 75-275 °C) to be clearly resolved in less than 2.20 × 104 μs using the IM(tof)MS instrument. Corresponding ions, masses, drift times, Ko values, and arbitrary signal intensities for each of the sample matrixes are reported for the CWA simulant DMMP. © 2005 American Chemical Society.

Cite

CITATION STYLE

APA

Steiner, W. E., Klopsch, S. J., English, W. A., Glowers, B. H., & Hill, H. H. (2005). Detection of a chemical warfare agent simulant in various aerosol matrixes by ion mobility time-of-flight mass spectrometry. Analytical Chemistry, 77(15), 4792–4799. https://doi.org/10.1021/ac050278f

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free