Abstract
The advent of accelerator-based mass spectrometry (AMS) radiocarbon dating has reduced the minimum sample size required to ca 1 mg from the 1 g of traditional beta counting techniques ( cf Wölfli, Polach & Andersen, 1984). However, it is clear that even smaller samples will be necessary for some fields of endeavor, particularly environmental work where the absolute quantities are small, perhaps of the order of a few micrograms (Currie et al , 1985). This raises serious problems in the handling and measurement of such small amounts, so that dilution will sometimes be required. Normally, sample dilution is accomplished by adding “dead” CO 2 to the combusted sample CO 2 , which requires very careful measurement of two gas pressures for the calculation of a dilution ratio. By forming a chemical derivative of a sample before combustion, gas pressure measurement is not necessary and an exact dilution ratio can be selected by judicous choice of the dilution reagent. This paper demonstrates that such a technique is possible for the AMS 14 C dating of derivatized amino acids.
Cite
CITATION STYLE
Gillespie, R. (1986). Chemical Isotope Dilution for 14 C Ams and the Potential for GC/AMS. Radiocarbon, 28(3), 1065–1067. https://doi.org/10.1017/s0033822200020129