A new quantitative reverse transcription (RT)-PCR assay for human immunodeficiency virus type 1 (HIV-1) RNA (Abbott LCx HIV RNA Quantitative assay) has been compared with the Organon NucliSens assay on 521 retrospective samples obtained from HIV-1-positive patients monitored during highly active antiretroviral therapy, 79 of whom were assayed also by the Chiron Quantiplex 3.0 system and on characterized panels. The LCx system showed a moderate correlation (r = 0.795) and gave higher results than the NucliSens system on 245 of 327 concordant positive samples, with similar sensitivity. Correlation with Quantiplex system results was higher (r = 0.943). LCx reproducibility was very good; the procedure was simple, well controlled, and rapid (up to 48 results in 7 h). The HIV RNA quantitative assay on the LCx system is suitable for routine use.
CITATION STYLE
Zanchetta, N., Nardi, G., Tocalli, L., Drago, L., Bossi, C., Pulvirenti, F. R., … Gismondo, M. R. (2000). Evaluation of the Abbott LCx HIV-1 RNA quantitative, a new assay for quantitative determination of human immunodeficiency virus type 1 RNA. Journal of Clinical Microbiology, 38(10), 3882–3886. https://doi.org/10.1128/jcm.38.10.3882-3886.2000