An investigation into the design of a device to treat haemorrhagic stroke

7Citations
Citations of this article
29Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

In this study, we present the design considerations of a device to assist in the potential treatment of hemorrhagic stroke with the aim of stopping blood from flowing out into brain tissue. We present and model three designs for the clinical scenarios when saccular aneurysms rupture in the middle cerebral artery in the brain. We evaluate and model these three designs using computer aided design software, SolidWorks, which allows the devices to be tested using finite element analysis and also enables us to justify that the materials chosen were suitable for potential use. Computational fluid dynamics modelling were used to demonstrate and analyse the flow of blood through the artery under conditions of normal and ruptured states. We conclude that our device could potentially be useful in the treatment of hemorrhagic stroke, and the modelling process is useful in assisting in determining the performance of our devices.

Cite

CITATION STYLE

APA

Randhawa, H. S., Pearce, G., Hepton, R., Wong, J., Zidane, I. F., & Ma, X. (2020). An investigation into the design of a device to treat haemorrhagic stroke. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 234(4), 323–336. https://doi.org/10.1177/0954411919890698