Abstract
BACKGROUND: Treatment with a cochlear implant (CI) poses the risk of inducing a behaviorally unmeasurable air-bone gap leading to false negative absence of cervical and ocular vestibular evoked myogenic potentials (cVEMPs, oVEMPs) to air conducted sound (ACS). OBJECTIVE: To investigate VEMP response rates to ACS and bone conducted vibration (BCV) in CI patients and the applicability of the B81 transducer for BCV stimulation. METHODS: Prospective experimental study including unilateral CI patients, measuring cVEMPs and oVEMPs to ACS and to BCV, comparing response rates, signed asymmetry ratios, latencies, and amplitudes. RESULTS: Data of 13 CI patients (mean age 44±12 years) were analyzed. For the CI side, oVEMP and cVEMP response rates were significantly higher for BCV (77%cVEMP, 62%oVEMP) compared to ACS (23%cVEMP, 8%oVEMP). For the contralateral side, no difference between response rates to ACS (85%cVEMP, 69%oVEMP) and BCV (85%cVEMP, 77%oVEMP) was observed. Substantially higher asymmetries were observed for ACS (-88±23%for cVEMPs, -96±11%for oVEMPs) compared to BCV (-12±45%for cVEMPs, 4±74%for oVEMPs). CONCLUSIONS: BCV is an effective stimulus for VEMP testing in CI patients. The B81 is a feasible stimulator.
Author supplied keywords
Cite
CITATION STYLE
Fröhlich, L., Wilke, M., Plontke, S. K., & Rahne, T. (2022). Bone conducted vibration is an effective stimulus for otolith testing in cochlear implant patients. Journal of Vestibular Research: Equilibrium and Orientation, 32(4), 355–365. https://doi.org/10.3233/VES-210028