Abstract
Objective: In the field of otorhinolaryngology, a variety of skull implants have been developed to assist hearing-impaired or even deaf patients. The first step in the implantation procedure, and also in lateral skull-base surgery, is to drill the calvarian or mastoid bone. We intended to investigate the hitherto unknown parameters for performing this procedure and to establish the first set-up for robotic milling of the lateral skull base. Materials and Methods: Experimental milling of the skull base was conducted on two human specimens using a hexapod robot. Results: Optimized parameters were determined with a drill speed of 30,000 revolutions/min and a form feed rate of 5 mm/s for the calvarium and 1 mm/s for mastoid bone, respectively, in a spiral-path fashion. While using a cutting burr, mean force levels were 4.81 N for calvarian bone and 6.12 N for mastoid bone, respectively-well below our empirical limit of 10 N. However, maximum levels easily surpassed these limits, reaching 27.7 N. Conclusion: The prerequisites for robotic skull-base surgery were fulfilled. With further work to implement feedback of sensory input, robots may increase precision for various tasks in skull-base surgery. © 2003 Informa UK Ltd All rights reserved.
Author supplied keywords
Cite
CITATION STYLE
Federspil, P. A., Plinkert, B., & Plinkert, P. K. (2003). Experimental robotic milling in skull-base surgery. Computer Aided Surgery, 8(1), 42–48. https://doi.org/10.3109/10929080309146102