Abstract
Purpose High-field magnetic resonance imaging (MRI) has the challenge of inhomogeneous B1+, and consequently inhomogeneous flip angle distribution, which causes spatially dependent contrast and makes clinical diagnosis difficult. Method We propose a two-step pulse design procedure in which (1) a combination of linear and nonlinear spatial encoding magnetic fields (SEMs) is used to remap the B1+ map in order to reduce the dimensionality of the problem, (2) the locations, amplitudes, and phases of spoke pulses are estimated in one dimension. The advantage of this B 1+ remapping is that when the isointensity contours of a linear combination of SEMs are similar to the isointensity contours of B 1+, a simple pulse sequence design using time-varying SEMs can achieve a homogenous flip-angle distribution efficiently. Results We demonstrate that spatially selective radiofrequency (RF) excitation with generalized SEMs (SAGS) using both linear and quadratic SEMs in a multi-spoke k-space trajectory can mitigate the B1+ inhomogeneity at 7T efficiently. Numerical simulations based on experimental data suggest that, compared with other methods, SAGS provide a formulation allowing multiple-pulse design, a similar average flip-angle distribution with less RF power, and/or a more homogeneous flip-angle distribution. Conclusion Without using multiple RF coils for parallel transmission, SAGS can be used to mitigate the B 1+ inhomogeneity in high-field MRI experiments. © 2013 Wiley Periodicals, Inc.
Author supplied keywords
Cite
CITATION STYLE
Hsu, Y. C., Chern, I. L., Zhao, W., Gagoski, B., Witzel, T., & Lin, F. H. (2014). Mitigate B1+ inhomogeneity using spatially selective radiofrequency excitation with generalized spatial encoding magnetic fields. Magnetic Resonance in Medicine, 71(4), 1458–1469. https://doi.org/10.1002/mrm.24801
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.