Abstract
We describe a new implantable hearing-aid output device, a piezoelectric bellows round-window driver (PBRD), which is expected to be unaffected by external magnetic fields. The core elements of the PBRD are a piezoelectric element and a gold-coated copper bellows of low stiffness that transmits piezoelectric displacements, without significant attenuation, to the round window (RW). We analyzed structural and mechanical vibrations when confirming bellows transmission efficiency using a finite element model simulation. The PBRD was bench-tested under no-load conditions to determine its frequency response characteristics. We compared the outputs of the PBRD and a commercial floating-mass transducer in situ in human cadaveric temporal bones with responses measured using a laser Doppler vibrometer. PBRD outputs were excellent at both low (0.1-0.7 kHz) and high (2-16 kHz) frequencies; thus, the PBRD has the potential to compensate for conductive and/or sensorineural hearing loss. The frequency-response performance of the PBRD is better than that of conventional RW drivers (actuators or transducers).
Author supplied keywords
Cite
CITATION STYLE
Shin, D. H., Seong, K. W., Nakajima, H. H., Puria, S., & Cho, J. H. (2020). A Piezoelectric Bellows Round-Window Driver (PBRD) for Middle-Ear Implants. IEEE Access, 8, 137947–137954. https://doi.org/10.1109/ACCESS.2020.3010539