Abstract
The feasibility of using the PRESAGE/Optical-CT system for 3D dosimetry verification around a brachytherapy source is investigated. Method and Materials: Brachytherapy dose distributions were obtained by irradiation of cylindrical PRESAGE volumes 6cm in diameter by 8cm height with a GammaMed 12i Ir-192 HDR unit (Varian Medical Systems). A narrow channel on the central axis was created by setting a steel catheter in the Presage during manufacture, enabling measurements close to the source (∼3mm). Results: Comparison of dose line profiles shows good agreement between PRESAGE and verified calculated dose calculation, in both high and low dose regions. Conclusion: The PRESAGE/Optical-CT shows good potential in verification of 3D dose distributions around brachytherapy sources. © 2010 IOP Publishing Ltd.
Cite
CITATION STYLE
Pierquet, M., Craciunescu, O., Steffey, B., Song, H., & Oldham, M. (2010). On the feasibility of verification of 3D dosimetry near brachytherapy sources using PRESAGE/optical-CT. In Journal of Physics: Conference Series (Vol. 250, pp. 452–456). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/250/1/012091