Abstract
Antithrombin (AT) plays a crucial role in the human anticoagulant system and has extensive clinical applications. However, traditional detection methods often require large sample volumes, complex procedures, and lengthy processing times. Methods: We integrated digital microfluidics technology with AT detection to develop a point-of-care testing (POCT) device that is user-friendly and fully automated for real-time AT testing. Results: This device allows for automation and enhanced adaptability to various settings, requiring only a minimal sample volume (whole capillary blood), thereby omitting steps such as plasma separation to save time and improve clinical testing efficiency. Comparisons with conventional AT activity detection methods demonstrate a high degree of consistency in the results obtained with this device. Conclusion: The AT detection system we developed exhibits significant effectiveness and holds substantial research potential, positioning it to evolve into a clinically impactful POCT solution for AT assessment.
Author supplied keywords
Cite
CITATION STYLE
Li, D., Hu, H., Zhang, H., Shang, L., Zhao, T., Zhao, Q., … Liu, X. (2025). Automated Antithrombin Activity Detection with Whole Capillary Blood Based on Digital Microfluidic Platform. Micromachines, 16(7). https://doi.org/10.3390/mi16070785