Abstract

The use of oral anticoagulation treatment (OAT) in patients with an international normalized ratio (INR) higher than 1.7 is a contraindication to thrombolysis in acute ischemic stroke. The aim of the present study is to compare the use of point-of-care (POC) coagulometers to the standard coagulation analysis (SCA) procedure of the INR as a decision-making test for use with patients taking OAT. Method: Eighty patients on chronic OAT underwent a POC and an SCA during a regular outpatient evaluation. Results: When comparing the abilities of the POC test and the SCA test to identify adequate levels for thrombolysis (≤1.7), the POC had a sensitivity of 96.6% (95%CI 88.4-99.1) and a specificity of 60.0% (95%CI 38.6-78). POC overestimated INR levels by 0.51 points compared to the SCA test. Conclusion: POC has a high sensitivity compared to the SCA test for the identification of patients within the cut-off point for thrombolysis.

Cite

CITATION STYLE

APA

Bruch, T. P., Mendes, D. C., Pedrozo, J. C., Figueiredo, L., Nóvak, E. M., Zétola, V. F., & Lange, M. C. (2014). Is point-of-care accurate for indicating thrombolysis in anticoagulated patients on oral anticoagulation treatments? Arquivos de Neuro-Psiquiatria, 72(7), 487–489. https://doi.org/10.1590/0004-282X20140075