The long readout times of single-shot acquisitions and the high field strengths desired for functional MRI (fMRI) using blood oxygenation level-dependent (BOLD) contrast make functional scans sensitive to magnetic field inhomogeneity. If it is not corrected during image reconstruction, field inhomogeneity can cause geometric distortions in the images when Cartesian k-space trajectories are used or blurring with spiral acquisitions. Many traditional methods to correct for field inhomogeneity distortions rely on a static field map measured with the use of images that are themselves distorted. In this work, we employ a regularized least-squares approach to jointly estimate both the undistorted image and field map at each acquisition using a spiral-in/spiral-out pulse sequence. Simulation and phantom studies show that this method is accurate and stable over a time series. Human functional studies show that the jointly estimated field map may be more accurate than standard field map estimates in the presence of respiration-induced phase oscillations, leading to better detection of functional activation. The proposed method measures a dynamic field map that accurately tracks magnetic field drift and respiration-induced phase oscillations during the course of a functional study. © 2004 Wiley-Liss, Inc.
CITATION STYLE
Sutton, B. P., Noll, D. C., & Fessler, J. A. (2004). Dynamic field map estimation using a spiral-in/spiral-out acquisition. Magnetic Resonance in Medicine, 51(6), 1194–1204. https://doi.org/10.1002/mrm.20079