Abstract
Purpose: To assess sound-level exposure during vitrectomy using 3 of the most common commercially available machines. Methods: This noninterventional cross-sectional study examined sound emission from the Constellation, Stellaris, and EVA vitrector systems. For each machine, a noise dosimeter was used to measure the sound-level exposure of the surgeon during 3 surgical cases in which vitrectomy was performed. Sound levels associated with progressively increasing cut rates and vacuum pressures were also measured. Finally, sound measurements were taken during the use of various additional functions of each machine, including diathermy, laser, and extrusion. Sound levels were compared with occupational health guidelines in Canada and the United States. Results: The maximum sound level recorded during vitrectomy surgery was 88.2 dBA. The mean sound level during vitrectomy surgical cases ranged from 58.5 to 66.8 dBA. A strong positive linear correlation was found between the cut rate and sound level (r = 0.88-0.98) and the vacuum pressure and sound level (r = 0.83-0.97). This relationship was consistent across the 3 vitrector systems (P
Author supplied keywords
Cite
CITATION STYLE
Ruparelia, S., Orr, S., Choudhry, N., Wong, R. W., Smith, C. A., Taylor, S. M., & Gupta, R. R. (2023). Risk for Surgical Team Hearing Loss With Vitrectomy. Journal of VitreoRetinal Diseases, 7(5), 397–403. https://doi.org/10.1177/24741264231172564