The direct and accurate determination of major elements Ca, K, Mg and Na in water by HR-ICPMS

17Citations
Citations of this article
43Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A direct, accurate and precise method is reported for major elements Ca, K, Mg and Na measurements in river and drinking water using a high resolution ICP-MS. Double isotope dilution calibration was used for the determination of Mg whereas the combined standard addition and internal standardization (Sc) was used for Ca, K and Na measurements. The method is validated by analysis fresh water SLRS-5, SLRS-6 and SRM1640a with satisfactory results characterized by high precisions of 0.055% to 0.66% RSD (or 0.29–1.8% relative combined uncertainty) for all four analytes, superior to those reported in earlier studies. As noted, use of internal standard Sc has significantly (3–33 times) improved measurement precisions for Ca, K and Na compared to standard addition calibration alone. The proposed method was applied for the determination of major elements Ca, K, Mg and Na in a candidate drinking water CRM AQUA-1. Values of 1.908 ± 0.007 µg g−1, 8.27 ± 0.14 µg g−1, 0.660 ± 0.010 µg g−1 and 13.76 ± 0.05 µg g−1 (u, k = 1) were obtained for Mg, Ca, K and Na in AQUA-1 drinking water, respectively.

Cite

CITATION STYLE

APA

Nadeau, K., Mester, Z., & Yang, L. (2018). The direct and accurate determination of major elements Ca, K, Mg and Na in water by HR-ICPMS. Scientific Reports, 8(1). https://doi.org/10.1038/s41598-018-34028-z

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free