Bead magnetorelaxometry with an on-chip magnetoresistive sensor

40Citations
Citations of this article
48Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Magnetorelaxometry measurements on suspensions of magnetic beads are demonstrated using a planar Hall effect sensor chip embedded in a microfluidic system. The alternating magnetic field used for magnetizing the beads is provided by the sensor bias current and the complex magnetic susceptibility spectra are recorded as the 2nd harmonic of the sensor response. The complex magnetic susceptibility signal appears when a magnetic bead suspension is injected, it scales with the bead concentration, and it follows the Cole-Cole expression for Brownian relaxation. The complex magnetic susceptibility signal resembles that from conventional magnetorelaxometry done on the same samples apart from an offset in Brownian relaxation frequency. The time dependence of the signal can be rationalized as originating from sedimented beads. © 2011 The Royal Society of Chemistry.

Cite

CITATION STYLE

APA

Dalslet, B. T., Damsgaard, C. D., Donolato, M., Strømme, M., Strömberg, M., Svedlindh, P., & Hansen, M. F. (2011). Bead magnetorelaxometry with an on-chip magnetoresistive sensor. Lab on a Chip, 11(2), 296–302. https://doi.org/10.1039/c0lc00002g