Cell density detector based on light beam focusing

2Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

Abstract

Although the lab-on-a-chip system has been successfully applied in a wide variety of fields, the goal of achieving a cell counter with simple operation, low cost, and high accuracy still attracts continuous research efforts. In this paper, the authors explore a cell counter based on light beam focusing to measure the density of adherent cells. In this sensor, the light emitted from the optical fibers is collimated by the collimating lens formed in polydimethylsiloxane (PDMS). The uniformly attached adherent cells act as a convex lens, focusing the collimated light propagated through them. The intensity of the focused light indicates the density of the adherent cells. For Hela cells, a detection limit of 8.3 × 104 cells/mL with a detection range from 0.1 × 106 cells/mL to 1.0 × 106 cells/mL is achieved. This sensor is particularly useful for drug screening, cell pathology analysis, and cancer pre-diagnosis.

Cite

CITATION STYLE

APA

Jian, A., Li, H., Zhang, Y., Duan, Q., Zhang, Q., & Sang, S. (2018). Cell density detector based on light beam focusing. Micromachines, 9(11). https://doi.org/10.3390/mi9110592