Effect of the detector width and gas pressure on the frequency response of a micromachined thermal accelerometer

10Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

Abstract

In the present work, the design and the environmental conditions of a micromachined thermal accelerometer, based on convection effect, are discussed and studied in order to understand the behavior of the frequency response evolution of the sensor. It has been theoretically and experimentally studied with different detector widths, pressure and gas nature. Although this type of sensor has already been intensively examined, little information concerning the frequency response modeling is currently available and very few experimental results about the frequency response are reported in the literature. In some particular conditions, our measurements show a cut-off frequency at -3dB greater than 200Hz. By using simple cylindrical and planar models of the thermal accelerometer and an equivalent electrical circuit, a good agreement with the experimental results has been demonstrated. © 2011 by the authors.

Cite

CITATION STYLE

APA

Garraud, A., Combette, P., Courteaud, J., & Giani, A. (2011). Effect of the detector width and gas pressure on the frequency response of a micromachined thermal accelerometer. Micromachines, 2(2), 167–178. https://doi.org/10.3390/mi2020167