This article describes a method that improves the performance of previous approaches for the automatic detection of the common carotid artery (CCA) lumen centerline on longitudinal B-mode ultrasound images. We propose to detect several lumen centerline candidates using local symmetry analysis based on local phase information of dark structures at an appropriate scale. These candidates are analyzed with selection mechanisms that use symmetry, contrast or intensity features in combination with position-based heuristics. Several experimental results are provided to evaluate the robustness and performance of the proposed method in comparison with previous approaches. These results lead to the conclusion that our proposal is robust to noise, lumen artifacts, contrast variations and that is able to deal with the presence of CCA-like structures, significantly improving the performance of our previous approach, from 87.5% ± 0.7% of correct detections to 98.3% ± 0.3% in a set of 200 images.
CITATION STYLE
Rouco, J., Azevedo, E., & Campilho, A. (2016). Automatic lumen detection on longitudinal ultrasound B-mode images of the carotid using phase symmetry. Sensors (Switzerland), 16(3). https://doi.org/10.3390/s16030350
Mendeley helps you to discover research relevant for your work.