Design and Simulation Study of a CNT-Based Multisource Cubical CT System for Dynamic Objects

6Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The purpose of this paper is to design and simulate a new computed tomography (CT) system with a high temporal resolution for dynamic objects. We propose a multisource cubical CT (MCCT) system with X-ray tubes and detectors installed on a cube. Carbon nanotube- (CNT-) based X-ray focal spots are distributed on the twelve edges of the cube. The distribution of X-ray focal spots and detectors completely avoids mechanical movements to scan an object under inspection. CNTs are excellent electron field emitters because the use of a "cold" cathode makes it possible to fabricate a cathode with multiple electron emission points, and the CNT-based X-ray focal spots possess little response time and programmable emission. The proposed rotation-free MCCT system can acquire a high scanning speed when using a high frame rate detector. A three-dimensional (3D) reconstruction algorithm with tensor framelet-based L0-norm (TF-L0) minimization is developed for the simulation study of the MCCT. Simulation experiment results demonstrate the feasibility of the MCCT system.

Cite

CITATION STYLE

APA

Gong, C., Zeng, L., Wang, C., & Ran, L. (2018). Design and Simulation Study of a CNT-Based Multisource Cubical CT System for Dynamic Objects. Scanning, 2018. https://doi.org/10.1155/2018/6985698