Multicentre evaluation of the Monarch (IL) clinical chemistry analyser

1Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A multicentre evaluation of the Monarch centrifugal analyser is reported. Precision, linearity and accuracy were assessed by comparison with routine methods. Calibration stability, photo-metric and dispensing accuracy, and carry-over related to samples and reagents were also evaluated. The overall performance of the instrument was good, showing an excellent photometric and dispensing accuracy, absence of sample-dependent carry-over, and almost negligible reagent carry-over. Good precision, linearity and correlation with routine methods were found for the parameters tested. The instrument is reliable and is now used as the routine clinical chemistry analyser in two of the three laboratories taking part in the evaluation. © 1988, Hindawi Publishing Corporation. All rights reserved.

Cite

CITATION STYLE

APA

Ceriotti, F., Bonini, P., Guagnellini, E., Okely, C., Zoppi, F., & Montalbetti, N. (1989). Multicentre evaluation of the Monarch (IL) clinical chemistry analyser. Journal of Automatic Chemistry, 11(5), 206–211. https://doi.org/10.1155/S1463924689000416