In vitro performance of Lifetech IBS Angel™ (iron-based bioresorbable scaffold) stents during overdilation for use in pediatric patients

2Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Objectives: The objective of this study was to assess the mechanical performance of the Lifetech IBS Angel stents during overdilation as is often required in pediatric applications; including time of first fracture, foreshortening, and the type of fracturing that occurs. Materials and methods: In vitro testing was performed and repeated for each stent three times under physiologic conditions with continuous audiovisual imaging allowing for post-testing evaluations. Assessment of sheath fit was also completed. Results: A total of 47 stents on monorail system were overdilated to complete fracture after passing through either a 4 or 5 French sheath. First strut fracture occurred in 4 and 6 mm stents when they reached greater than 50% overexpansion. Larger stents could achieve at least 30% increased diameter prior to first strut fracture. No fragmentation of any of the stents was seen throughout testing. Conclusion: The IBS Angel has thin struts allowing for a lower profile with increased maneuverability and use with smaller sheaths. Embolization potential of strut fragments was not seen. Increased diameter well beyond design parameters was seen in all with acceptable foreshortening.

Cite

CITATION STYLE

APA

Bjorkman, K., Maldonado, J. R., Saey, S., & McLennan, D. (2022). In vitro performance of Lifetech IBS AngelTM (iron-based bioresorbable scaffold) stents during overdilation for use in pediatric patients. Frontiers in Cardiovascular Medicine, 9. https://doi.org/10.3389/fcvm.2022.1006063

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free