Abstract
The operational experience with a highly-granular analogue hadronic calorimeter (AHCAL) consisting of 7608 individual scintillator tiles readout via Silicon-Photo-multipliers (SiPM) is presented. The calibration of each cell is based on minimum ionizing particle signals for which in general a muon beam is used. In addition, a correction for the non-linearity introduced by the finite number of pixels (1156) in the SiPM is applied. The aspects of temperature and voltage dependence of SiPM are addressed, and monitoring and calibration procedures are discussed. Such procedures are essential for the extrapolation of calibration factors over several days of data taking with the calorimeter. For this purpose a versatile UV-LED light distribution system was developed, capable of delivering light to all tiles with intensity from a few photo-electrons to the saturation of the SiPM. The procedures are tested using data collected with the AHCAL at the CERN SPS test beam. © 2009 IOP Publishing Ltd.
Cite
CITATION STYLE
Lucaci-Timoce, A. (2009). Calibration and monitoring of a scintillator HCAL with SiPMs CALICE scintillator HCAL. Journal of Physics: Conference Series, 160. https://doi.org/10.1088/1742-6596/160/1/012027
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.