Abstract
Purpose: To compare retinal thickness (RT) measurements obtained by spectral-domain optical coherence tomography (SD-OCT) and time-domain OCT (TD-OCT) in eyes with and without posterior segment disease diagnoses. Design: Retrospective, observational chart review. Methods: Scans of 184 healthy and diseased eyes from 106 patients were included in the study. Pair-wise comparisons of RT measurements from SD-OCT machines from 2 different manufacturers and TD-OCT were performed for all scans and with eyes stratified by disease diagnosis. Results: Foveal thickness measurements obtained from both SD-OCT machines were greater than those measured by TD-OCT, by 51.0 ± 23.8 μm and 72.5 ± 30.4 μm, respectively. The foveal thickness difference between the 2 SD-OCT models was 24.3 ± 19.1 μm. Foveal thickness differences varied by disease diagnosis. Conclusions: RT measurements obtained by SD-OCT are consistently greater than those obtained by TD-OCT. RT measurement differences may vary by disease diagnosis or by SD-OCT model. Consideration of these measurement differences is essential when OCT-determined RT measurement data are used in clinical settings. © 2009 Elsevier Inc. All rights reserved.
Cite
CITATION STYLE
Han, I. C., & Jaffe, G. J. (2009). Comparison of Spectral- and Time-Domain Optical Coherence Tomography for Retinal Thickness Measurements in Healthy and Diseased Eyes. American Journal of Ophthalmology, 147(5). https://doi.org/10.1016/j.ajo.2008.11.019