Abstract
We present an open-source, picture archiving and communication system (PACS)-integrated radiation exposure extraction engine (RE3) that provides study-, series-, and slice-specific data for automated monitoring of computed tomography (CT) radiation exposure. RE3 was built using open-source components and seamlessly integrates with the PACS. RE3 calculations of dose length product (DLP) from the Digital imaging and communications in medicine (DICOM) headers showed high agreement (R2 = 0.99) with the vendor dose pages. For study-specific outlier detection, RE3 constructs robust, automatically updating multivariable regression models to predict DLP in the context of patient gender and age, scan length, water-equivalent diameter (Dw), and scanned body volume (SBV). As proof of concept, the model was trained on 811 CT chest, abdomen + pelvis (CAP) exams and 29 outliers were detected. The continuous variables used in the outlier detection model were scan length (R2 = 0.45), Dw (R2 = 0.70), SBV (R2 = 0.80), and age (R2 = 0.01). The categorical variables were gender (male average 1182.7 ± 26.3 and female 1047.1 ± 26.9 mGy cm) and pediatric status (pediatric average 710.7 ± 73.6 mGy cm and adult 1134.5 ± 19.3 mGy cm).
Author supplied keywords
Cite
CITATION STYLE
Weisenthal, S. J., Folio, L., Kovacs, W., Seff, A., Derderian, V., Summers, R. M., & Yao, J. (2016). Open-Source Radiation Exposure Extraction Engine (RE3) with Patient-Specific Outlier Detection. Journal of Digital Imaging, 29(4), 406–419. https://doi.org/10.1007/s10278-015-9852-y