Abstract

A superconducting self-shielding three-solenoid system with an adjustable shielding factor is developed, implemented, and characterized using a single antiproton in a Penning trap. With the tuned system, we suppress external magnetic field disturbances by up to a factor of 225 ± 15, allowing antiproton-to-proton charge-to-mass ratio comparisons with fourfold reduced frequency fluctuations and antiproton magnetic moment determinations with tenfold reduced uncertainty.

Cite

CITATION STYLE

APA

Devlin, J. A., Wursten, E., Harrington, J. A., Higuchi, T., Blessing, P. E., Borchert, M. J., … Ulmer, S. (2019). Superconducting Solenoid System with Adjustable Shielding Factor for Precision Measurements of the Properties of the Antiproton. Physical Review Applied, 12(4). https://doi.org/10.1103/PhysRevApplied.12.044012