Permeability and secondary membrane formation of a high flux polysulfone hemofilter

129Citations
Citations of this article
32Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

It has been assumed that the molecular weight (MW) cut-off of a newly fabricated polysulfone capillary dialyzer (F60, Fresenius, FRG) is similar to that of the human glomerulus. We recently tested the device in vivo and found this not to be so, based on the device's ability to eliminate substances of a MW of 10,000 to 60,000 daltons. One of the reasons for this discrepancy was found to be the influence of secondary membrane formation on solute permeability. Endogenous marker substances of a defined MW (β2-microglobulin, myoglobin, RBP, α1-microglobulin, acid α1-glycoprotein, α1-antitrypsin, prealbumin, and albumin were measured by laser nephelometry or radioimmune assay; sieving coefficients (SC) and protein eliminations were calculated for each low MW protein.

Cite

CITATION STYLE

APA

Rockel, A., Hertel, J., Fiegel, P., Abdelhamid, S., Panitz, N., & Walb, D. (1986). Permeability and secondary membrane formation of a high flux polysulfone hemofilter. Kidney International, 30(3), 429–432. https://doi.org/10.1038/ki.1986.202