Abstract

We have studied response times of 30 anaesthetists to a standardized episode of arterial oxygen desaturation in a simulated patient, randomized to the use of either a fixed or variable pitch pulse oximeter. We wished to determine if a variable auditory signal was important in detecting adverse events. A variable pitch pulse signal had a shorter time to recognition of desaturation (P < 0.0001), with a mean response time of 32 s, compared with 129 s for the fixed pitch signal.

Cite

CITATION STYLE

APA

Craven, R. M., & McIndoe, A. K. (1999). Continuous auditory monitoring - How much information do we register? British Journal of Anaesthesia, 83(5), 747–749. https://doi.org/10.1093/bja/83.5.747