Fourier transform spectrometer for spectral emissivity measurement in the temperature range between 60 and 1500°C

52Citations
Citations of this article
24Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A new spectral emissivity measurement system has been developed at Harbin Institute of Technology (HIT) by using a Fourier transform infrared (FTIR) spectrometer. The spectral range between 0.6 and 25 νm was covered by a photovoltaic HgCdTe and a silicon photodiode detector. A SiC heater with a black hole was employed for heating the sample. The temperature of the sample can be controlled in a range between 60 and 1500°C with an error of less than 1°C. The system was calibrated against two high quality reference blackbodies: a low temperature heat-pipe blackbody operated in the temperature range between 60°C and 300°C and a high temperature blackbody with SiC heater operated in the temperature range between 300°C and 1500°C. Several tests were done for this new system. The estimated uncertainty of emissivity measurement is better than 3%. © 2005 IOP Publishing Ltd.

Cite

CITATION STYLE

APA

Dai, J., Wang, X., & Yuan, G. (2005). Fourier transform spectrometer for spectral emissivity measurement in the temperature range between 60 and 1500°C. In Journal of Physics: Conference Series (Vol. 13, pp. 63–66). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/13/1/015