Early structural valve deterioration of balloon expandable transcatheter aortic valve leaflets due to intrinsic and extrinsic nodular calcifications in a haemodialysis patient: a case report

1Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: Several reports have shown that transcatheter aortic valves are comparable in durability to surgical aortic valves. However, early structural valve deterioration (SVD) is rarely reported to occur, especially in haemodialysis patients. Case summary: We present a case of rapidly progressive bioprosthetic aortic valve stenosis in a patient with end-stage renal disease secondary to diabetic nephropathy in an 83-year-old female admitted due to progressive dyspnoea and orthopnoea. A 23 mm sized SAPIEN3 bioprosthetic aortic valve showed normal function for the first year after transcatheter aortic valve implantation (TAVI), but then rapidly developed stenosis and required acute hospitalization for heart failure a year and a half after surgery. Emergent surgical aortic valve replacement with a 19 mm On-X valve (CryoLife, Kennesaw, GA, USA) was performed. Pathological examination of the explanted SAPIEN 3 valve demonstrated severely degenerated bioprosthetic pericardial leaflets with severe intrinsic and extrinsic nodular calcifications, which could limit the leaflet motion. Discussion: There is a lack of reports on the long-term procedural outcomes of TAVI in haemodialysis patients. The development of SVD in patients undergoing dialysis is multifactorial and has yet to be fully elucidated. In the presented case, the removed TAVI valve had severe extrinsic calcified nodules alongside a fibrin thrombus. Considering these pathological findings, antithrombotic therapy to prevent fibrin thrombus from adhering to the TAVI valve may be important to avoid early SVD.

Cite

CITATION STYLE

APA

Onishi, K., Mizutani, K., Sato, Y., & Nakazawa, G. (2024). Early structural valve deterioration of balloon expandable transcatheter aortic valve leaflets due to intrinsic and extrinsic nodular calcifications in a haemodialysis patient: a case report. European Heart Journal - Case Reports, 8(6). https://doi.org/10.1093/ehjcr/ytae265

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free